Fits Tiger 320 Series

RTD & SMART 16-BIT LOAD CELL INPUT MODULE

State-of-the-art Electromagnetic Noise

Suppression Circuitry.

Ensures signal integrity even in harsh EMC environments.

Constant Current Source.

Accurate

RTD excitation.

11-pin Input Connector.

Load cell and RTD sensors.

On-board Excitation Voltage

Generator.

+5 V, 130 mA.

Crystal Controlled 50/60 Hz Line Frequency Rejection.

Load Cell Bridge Header.

6-wire or 4-wire Selectable.

Ultra Low-noise 16-bit

A/D Converter.

Approaching 19-bit performance with additional software filtering.

Interface to Tiger Meter.

On-board Digital Signal Processor.

Provides 6 output functions.

RTD & high performance load cell combination.

A smart input module designed to monitor temperature using an RTD sensor and weight using a load cell.

Ideal for applications that calculate fluid volume from weight and density measurements corrected for temperature changes.

Combined with the powerful Tiger 320 Series programmable meter controller, continuous monitoring and control of accurate volume versus cost calculations are a reality.

Input Module

Order Code Suffix

ISSB (50/60 Hz)

RTD (CH1)

Excitation

Sensor Types

Wiring

Hardware Module Specifications

1.5 mA DC continuous.

Pt (100 ) 385 & 395, Ni 120 & Cu 10.

3-wire excitation & lead compensation.

(CH2 / CH3 / CH4)

Load Cell

Excitation

Input Range

Input Sensitivity

Zero Drift

Span Drift

Non-linearity

Input noise

Signal Processing Rate

Wiring

Guard

5 V DC, 130 mA maximum.

Software selectable for sensors 1 mV/V to 20 mV/V.

0.08 V/ count maximum. 40 nV/C typical. 5 ppm/C of full scale maximum. 0.003% of full scale maximum.

160 nV pp typical at 1 Hz output rate.

50 Hz maximum, 1 Hz minimum.

4-wire & 6-wire selectable.

Shield 2.5 V common mode voltage.

Load Cell

Dual Output Rates

Peak & Valley Outputs Capture Output

Rate of Change Output Line Frequency Rejection

Software Module Features

Rapid and average response outputs.

Monitoring over and under-shoots.

Hardwire signal capture.

Useful for fine tuning reaction times.

50/60 Hz selectable.

ISSB Data Sheet (NZ347)

Texmate, Inc. Tel. (760) 598-9899 www.texmate.com

Page 1

Connector Pinouts

Figure 1 ISSB Input Module Showing 4-wire Load Cell

Figure 2 ISSB Input Module Showing 6-wire Bridge and

and 3-wire RTD Configuration

3-wire RTD Configuration

Smart Setup Registers

The meter uses three smar t setup registers to configure all smar t input modules. The ISSB input module requires only smart register 1 to be set up.

Input

Signal

Figure 3 ISSB Smar t Setup Registers Operational Flow Diagram

Internally, the RTD input is automatically assigned to channel 1

(CH1) and must be set up through Code 2. The RTD input feeds directly through the input module m ultiplexer to the meter s 17bit A/D converter, and then to the microprocessor for further processing.

The load cell input is processed in the input module s 16-bit A/D converter and digital signal processor. It is then f ed to some or all of the three remaining channels, CH2, CH3, CH4, where one of six outputs can be selected f or each channel from the output register map. Smart register 1 must be set up for the load cell to define the line frequency rejection, the sensor input in milliv olts per volt, and the output r ate.

Figure 4 ISSB Smart Setup Registers Signal Flow Diagram

Programming Quick Start Guide

Load Cell Setup

1 Press the

and

buttons at the same time to enter the main prog ramming mode.

Press the

button three times to enter Code 2. Set Code 2 to [X77].

This setting enters the smart register 1 code setup menu of the load cell input.

Note the output register map is different for each smart input module.

Page 2

Texmate, Inc. Tel. (760) 598-9899 www.texmate.com

ISSB Data Sheet (NZ347)

Press the

button.

This men u pro vides settings unique to smart register 1 of input module ISSB.

Using the

cell input, and output rate settings.

buttons, select the relevant line frequency rejection, load

5 Press the

button. The display returns to [Cod_2] [X77].

RTD Setup

6 Using the

buttons, set Code 2 to [X2X] to select a 3-wire RTD sensor for the CH1 measurement task.

In the 3rd digit, select the R TD type.

Press the

button to save the settings.

Press the

settings and return to the operational display.

and

buttons at the same time to save the RTD Code 2

Select a Load Cell Channel

Select the load cell output register f or CH2, CH3, or CH4

8 Press the

and

button at the same time again to re-enter the main prog ramming mode.

Press the

button five times to enter Code 4.

To select an output register for CH2, enter Code 4 and select the required register map settings f or CH2 in the 2nd digit.

Note, the 1st and 3rd digits must be set to 0.

CH2

Note the capture and rate of change registers can not be selected in Code 4 for CH2.

If required enter Code 5 and select the required post processing setting f or CH3 in the 2nd digit and the required output register map setting in the 3rd digit.

CH3

Note the register map is different for each smart input module type.

ISSB Data Sheet (NZ347)

Texmate, Inc. Tel. (760) 598-9899 www.texmate.com

Page 3

If required enter Code 6 and select the required post processing setting f or CH4 in the 1st digit and the required register map setting in the 3rd digit.

CH4

Press the

button to save the settings.

Press the

time to return to the operational display.

buttons at the same

and

Figure 5 Example Milk Dispensing Tank

Customer Configuration Settings:

RTD only

CH1

Load Cell only

1st Digit

2nd Digit 3rd Digit

1st Digit

2nd Digit 3rd Digit

1st Digit

2nd Digit 3rd Digit

CH2

CH3

CH4

1st Digit

2nd Digit 3rd Digit

1st Digit

2nd Digit 3rd Digit

1st Digit

2nd Digit 3rd Digit

For product details visit www.texmate.com

Tel: 1-760-598-9899 USA 1-800-839-6283 Thats 1-800-TEXMATE

Page 4

Texmate, Inc. Tel. (760) 598-9899 www.texmate.com

ISSB Data Sheet (NZ347)