State-of-the-art Electromagnetic Noise
Suppression Circuitry.
Ensures signal integrity even in harsh EMC environments.
Constant Current Source.
Accurate
RTD excitation.
11-pin Input Connector.
Load cell and RTD sensors.
On-board Excitation Voltage
Generator.
+5 V, 130 mA.
Crystal Controlled 50/60 Hz Line Frequency Rejection.
Load Cell Bridge Header.
6-wire or 4-wire Selectable.
Ultra Low-noise 16-bit
A/D Converter.
Approaching 19-bit performance with additional software filtering.
Interface to Tiger Meter.
On-board Digital Signal Processor.
Provides 6 output functions.
RTD & high performance load cell combination.
A smart input module designed to monitor temperature using an RTD sensor and weight using a load cell.
Ideal for applications that calculate fluid volume from weight and density measurements corrected for temperature changes.
Combined with the powerful Tiger 320 Series programmable meter controller, continuous monitoring and control of accurate volume versus cost calculations are a reality.
Input Module
Order Code Suffix
ISSB (50/60 Hz)
RTD (CH1)
Excitation
Sensor Types
Wiring
Hardware Module Specifications
1.5 mA DC continuous.
Pt (100 ) 385 & 395, Ni 120 & Cu 10.
3-wire excitation & lead compensation.
(CH2 / CH3 / CH4)
Load Cell
Excitation
Input Range
Input Sensitivity
Zero Drift
Span Drift
Non-linearity
Input noise
Signal Processing Rate
Wiring
Guard
5 V DC, 130 mA maximum.
Software selectable for sensors 1 mV/V to 20 mV/V.
0.08 V/ count maximum. 40 nV/C typical. 5 ppm/C of full scale maximum. 0.003% of full scale maximum.
160 nV pp typical at 1 Hz output rate.
50 Hz maximum, 1 Hz minimum.
4-wire & 6-wire selectable.
Shield 2.5 V common mode voltage.
Load Cell
Dual Output Rates
Peak & Valley Outputs Capture Output
Rate of Change Output Line Frequency Rejection
Software Module Features
Rapid and average response outputs.
Monitoring over and under-shoots.
Hardwire signal capture.
Useful for fine tuning reaction times.
50/60 Hz selectable.
ISSB Data Sheet (NZ347)
Texmate, Inc. Tel. (760) 598-9899 www.texmate.com
Page 1
Figure 1 ISSB Input Module Showing 4-wire Load Cell
Figure 2 ISSB Input Module Showing 6-wire Bridge and
and 3-wire RTD Configuration
3-wire RTD Configuration
The meter uses three smar t setup registers to configure all smar t input modules. The ISSB input module requires only smart register 1 to be set up.
Input
Signal
Figure 3 ISSB Smar t Setup Registers Operational Flow Diagram
Internally, the RTD input is automatically assigned to channel 1
(CH1) and must be set up through Code 2. The RTD input feeds directly through the input module m ultiplexer to the meter s 17bit A/D converter, and then to the microprocessor for further processing.
The load cell input is processed in the input module s 16-bit A/D converter and digital signal processor. It is then f ed to some or all of the three remaining channels, CH2, CH3, CH4, where one of six outputs can be selected f or each channel from the output register map. Smart register 1 must be set up for the load cell to define the line frequency rejection, the sensor input in milliv olts per volt, and the output r ate.
Figure 4 ISSB Smart Setup Registers Signal Flow Diagram
1 Press the
and
buttons at the same time to enter the main prog ramming mode.
Press the
button three times to enter Code 2. Set Code 2 to [X77].
This setting enters the smart register 1 code setup menu of the load cell input.
Note the output register map is different for each smart input module.
Page 2
Texmate, Inc. Tel. (760) 598-9899 www.texmate.com
ISSB Data Sheet (NZ347)
Press the
button.
This men u pro vides settings unique to smart register 1 of input module ISSB.
Using the
cell input, and output rate settings.
buttons, select the relevant line frequency rejection, load
5 Press the
button. The display returns to [Cod_2] [X77].
6 Using the
buttons, set Code 2 to [X2X] to select a 3-wire RTD sensor for the CH1 measurement task.
In the 3rd digit, select the R TD type.
Press the
button to save the settings.
Press the
settings and return to the operational display.
and
buttons at the same time to save the RTD Code 2
Select the load cell output register f or CH2, CH3, or CH4
8 Press the
and
button at the same time again to re-enter the main prog ramming mode.
Press the
button five times to enter Code 4.
To select an output register for CH2, enter Code 4 and select the required register map settings f or CH2 in the 2nd digit.
Note, the 1st and 3rd digits must be set to 0.
Note the capture and rate of change registers can not be selected in Code 4 for CH2.
If required enter Code 5 and select the required post processing setting f or CH3 in the 2nd digit and the required output register map setting in the 3rd digit.
Note the register map is different for each smart input module type.
ISSB Data Sheet (NZ347)
Texmate, Inc. Tel. (760) 598-9899 www.texmate.com
Page 3
If required enter Code 6 and select the required post processing setting f or CH4 in the 1st digit and the required register map setting in the 3rd digit.
Press the
button to save the settings.
Press the
time to return to the operational display.
buttons at the same
and
Figure 5 Example Milk Dispensing Tank
RTD only
Load Cell only
1st Digit
2nd Digit 3rd Digit
1st Digit
2nd Digit 3rd Digit
1st Digit
2nd Digit 3rd Digit
1st Digit
2nd Digit 3rd Digit
1st Digit
2nd Digit 3rd Digit
1st Digit
2nd Digit 3rd Digit
For product details visit www.texmate.com
Tel: 1-760-598-9899 USA 1-800-839-6283 Thats 1-800-TEXMATE
Page 4
Texmate, Inc. Tel. (760) 598-9899 www.texmate.com
ISSB Data Sheet (NZ347)
Texmate is an American manufacturer of high durability industrial grade panel meters, bar graphs, and transducers. Texmate's meters are frequently equipped with relays and various signal outputs for industrial automation applications and are known for their extremely long service life.